U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIDENT 10° X3 INSERT 36MM ID; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIDENT 10° X3 INSERT 36MM ID; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 623-10-36G
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/17/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate that devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Not available.
 
Event or Problem Description
Surgeon tried to fit a trident 10 degree liner (x3) into trident cup.Would not sit, regardless of troubleshooting.New liner sat perfectly first time.
 
Event or Problem Description
Surgeon tried to fit a trident 10 degree liner (x3) into trident cup.Would not sit, regardless of troubleshooting.New liner sat perfectly first time.
 
Additional Manufacturer Narrative
An event regarding seating/locking issue involving a trident liner was reported.The event was not confirmed.Method & results: -device evaluation and results: visual inspection visual inspection was performed as part of the material evaluation and indicated the following comments: damage observed on insert consistent with attempted implantation dimensional inspection: not performed as the reported device was implanted and subsequently explanted.The device was returned damaged and in its current condition would not be an accurate reflection of its original manufactured condition.Functional inspection: not performed as the reported device was implanted and subsequently explanted.The device was returned damaged and in its current condition would not be an accurate reflection of its original manufactured condition.Material analysis: material evaluation was completed and indicated the following comments: ma: damage observed on insert consistent with attempted implantation.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined.- medical records received and evaluation: no medical records were received for review with a clinical consultant.- device history review: all devices in the reported lot were manufactured and accepted into final stock with no relevant reported discrepancies.- complaint history review: there have been no other similar events for the reported lot.Conclusions: material evaluation was completed and indicated the following comments: damage observed on insert consistent with attempted implantation.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined.However, the exact cause of the event cannot be determined based on the information provided.Further information such as operative report are needed to complete the investigation for determining root cause.No further investigation is possible at this time as insufficient information was received.If additional information and/or the device becomes available the investigation will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT 10° X3 INSERT 36MM ID
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key8245315
Report Number0002249697-2019-00169
Device Sequence Number16695156
Product Code LPH
UDI-Device Identifier07613327039931
UDI-Public07613327039931
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
K033716
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2022
Device Catalogue Number623-10-36G
Device Lot NumberJ0332P
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/11/2019
Initial Date Received by Manufacturer 12/17/2018
Supplement Date Received by Manufacturer04/02/2019
Initial Report FDA Received Date01/14/2019
Supplement Report FDA Received Date04/29/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-