• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. FMP HIP; FMP HEMISPHERICAL SHELLS W/SCREW HOLES, 50MM, W/P2 COATING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. FMP HIP; FMP HEMISPHERICAL SHELLS W/SCREW HOLES, 50MM, W/P2 COATING Back to Search Results
Model Number 430-98-050
Health Effect - Clinical Code Pain (1994)
Date of Event 12/20/2018
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - the patient had impingement superiority and pain.The x-rays looked fine, all components were well seated and were not loose at the time of the revision.The surgeon chose to explant the acetabular cup, poly, and screw, and revise to a biomet g7 dual mobility.
 
Additional Manufacturer Narrative
The reason for this revision surgery was pain.The previous surgery and the surgery detailed in this investigation occurred 3.8 months apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.This investigation is limited in scope as the item(s) associated with this investigation was not returned to djo surgical - austin for examination.A review of the implant device history records (dhr), shows that the reported component(s) used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports associated with the product(s) that may have contributed to the reported event.The device(s) was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device(s) showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to pain.No other information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.There were no findings during this investigation that indicate that the reported device(s) was defective.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FMP HIP
Common Device Name
FMP HEMISPHERICAL SHELLS W/SCREW HOLES, 50MM, W/P2 COATING
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key8245631
Report Number1644408-2019-00010
Device Sequence Number16760176
Product Code LPH
UDI-Device Identifier00888912084017
UDI-Public(01)00888912084017
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
PMA/510(K) Number
K974093
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/23/2024
Device Model Number430-98-050
Device Catalogue Number430-98-050
Device Lot Number588G1212
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/20/2018
Supplement Date Received by Manufacturer02/05/2019
Initial Report FDA Received Date01/14/2019
Supplement Report FDA Received Date02/21/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
010-55-040, LOT 010A1149; 400-03-361, LOT 864B1269; 931-36-750, LOT 010V1031; 010-55-040, LOT 010A1149; 400-03-361, LOT 864B1269; 931-36-750, LOT 010V1031
Outcome Attributed to Adverse Event Required Intervention;
-
-