• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US MENISCAL DEPLOYMENT GUN; ORTHOPAEDIC CERCLAGE APPLIER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US MENISCAL DEPLOYMENT GUN; ORTHOPAEDIC CERCLAGE APPLIER Back to Search Results
Catalog Number 228143
Device Problem Firing Problem (4011)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/17/2016
Event Type  malfunction  
Manufacturer Narrative
If additional information should become available, a supplemental medwatch will be submitted accordingly.Udi: (b)(4).The lot number is currently unavailable; therefore, the exp date is unavailable.The complaint device is not being returned, therefore unavailable for a physical evaluation.A review into the depuy synthes mitek complaints system revealed no other complaints for this lot of devices that were released to distribution.We cannot discern a root cause for the reported failure mode.At this point in time, no corrective action is required and no further action is warranted.This file will remain receptive to any potential forthcoming information received that is pertinent and germane to this issue.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.A non-conformance search was performed for this product code 228143, lot 3860947 combination and no non-conformances were identified.No further information regarding the technique or instruments used has been provided to determine a root cause for this failure.If any additional information is obtained, this complaint will be re-opened to capture that information.At this point in time, no corrective action is required, and no further action is warranted.This report is being filed from the etq complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
 
Event Description
This is report 1 of 2 for the same event.It was reported by the sales rep that during a meniscal repair, it was observed that the coil got stuck when deploying the 2nd implant of the customer's omnispan meniscal repair system with 12 degree needle.The sales rep reported that the coil went under the 2nd implant, got stuck and did not advance it.The surgeon cut the suture between them and left the 1st implant in the meniscus.The surgeon completed the procedure with another like omnispan meniscal applier and another same like implant right adjacent to the original implant.There were no patient consequences or delays reported.The sales rep reported that the surgeon used two more implants to complete the repair and did not feel that the first implant failing caused the plan to be altered.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENISCAL DEPLOYMENT GUN
Type of Device
ORTHOPAEDIC CERCLAGE APPLIER
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6013142063
MDR Report Key8245700
MDR Text Key133846857
Report Number1221934-2019-55995
Device Sequence Number1
Product Code GEF
UDI-Device Identifier10886705010059
UDI-Public10886705010059
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/17/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number228143
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/17/2016
Initial Date FDA Received01/14/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-