If additional information should become available, a supplemental medwatch will be submitted accordingly.Udi: (b)(4).The lot number is currently unavailable; therefore, the exp date is unavailable.The complaint device is not being returned, therefore unavailable for a physical evaluation.A review into the depuy synthes mitek complaints system revealed no other complaints for this lot of devices that were released to distribution.We cannot discern a root cause for the reported failure mode.At this point in time, no corrective action is required and no further action is warranted.This file will remain receptive to any potential forthcoming information received that is pertinent and germane to this issue.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.A non-conformance search was performed for this product code 228143, lot 3860947 combination and no non-conformances were identified.No further information regarding the technique or instruments used has been provided to determine a root cause for this failure.If any additional information is obtained, this complaint will be re-opened to capture that information.At this point in time, no corrective action is required, and no further action is warranted.This report is being filed from the etq complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
|
This is report 1 of 2 for the same event.It was reported by the sales rep that during a meniscal repair, it was observed that the coil got stuck when deploying the 2nd implant of the customer's omnispan meniscal repair system with 12 degree needle.The sales rep reported that the coil went under the 2nd implant, got stuck and did not advance it.The surgeon cut the suture between them and left the 1st implant in the meniscus.The surgeon completed the procedure with another like omnispan meniscal applier and another same like implant right adjacent to the original implant.There were no patient consequences or delays reported.The sales rep reported that the surgeon used two more implants to complete the repair and did not feel that the first implant failing caused the plan to be altered.
|