COVIDIEN MFG DC BOULDER RAPIDVAC; APPARATUS, EXHAUST, SURGICAL Back to Search Results
Model Number SE3695
Medical Device Problem Code Self-Activation or Keying (1557)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 12/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation of the reported condition was performed.A complaint of the device maintaining the suction permanently was reported.The suction should be activated only at the moment when a diathermal burn is performed.One device was received as a service request and a visual inspection and functional test were performed by the field service team.A dhr review was completed for serial# (b)(4).There were no manufacturing issues related to the complaint for this serial number.According to the technician, the unit works as intended.Additionally, the motor was found to be burned and needed replacement.The firmware was already updated to 1.18.The device as received met the specifications and passed functional testing for the suction capability.It failed for the motor component.This complaint will be considered as unconfirmed for issues with the suction.A root cause for the motor issue is wear and tear.The unit has exceeded its useful life.As part of continuous improvement efforts and a corrective action, the motor was replaced.The device was fully checked for functionality and patient electrical safety according to norm 60601 cl-1 and the manufacturer's guideline.It was deemed technically efficient and safe to use.The unit was cleaned and disinfected according to the manufacturer's guideline.If additional information is received, the investigation will resume as needed.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during testing, when the forceps button on the front panel is chosen the device maintains the suction permanently.The suction should be activated only at the moment when a diathermal burn is performed.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAPIDVAC
Common Device Name
APPARATUS, EXHAUST, SURGICAL
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow drive
boulder CO 80301
MDR Report Key8246186
Report Number1717344-2019-00090
Device Sequence Number2203285
Product Code FYD
UDI-Device Identifier10884524001753
UDI-Public10884524001753
Combination Product (Y/N)N
Initial Reporter CountryFI
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,foreign,health profe
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 01/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSE3695
Device Catalogue NumberSE3695
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/17/2018
Initial Report FDA Received Date01/14/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-