• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE FLOWPORT II CANNULA WITH OBTURATOR STRYKER; ACCESSORIES, ARTHROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE FLOWPORT II CANNULA WITH OBTURATOR STRYKER; ACCESSORIES, ARTHROSCOPIC Back to Search Results
Catalog Number CAT01548
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/19/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device manufacture date is not known at this time.However, should it become available it will be provided in future reports.Additional information will be provided once the investigation has been completed.
 
Event or Problem Description
It was reported that the device broke during surgery.All pieces were retrieved and the procedure was completed successfully.
 
Additional Manufacturer Narrative
The product was not returned for investigation therefore the reported failure mode was not confirmed.Alleged failure: break.Probable root cause: design - design stackup interference - insufficient assembly design concept manufacturing - cannula, instrument or scope not manufactured to specification application - excessive force - user unfamiliarity with devices.The reported failure mode will be monitored for future reoccurrence.The device manufacture date is not known.
 
Event or Problem Description
It was reported that the device broke during surgery.All pieces were retrieved and the procedure was completed successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLOWPORT II CANNULA WITH OBTURATOR STRYKER
Common Device Name
ACCESSORIES, ARTHROSCOPIC
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key8248405
Report Number0002936485-2019-00020
Device Sequence Number2462916
Product Code NBH
UDI-Device Identifier07613252633037
UDI-Public07613252633037
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberCAT01548
Device Lot Number003147
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/19/2018
Supplement Date Received by Manufacturer12/19/2018
Initial Report FDA Received Date01/15/2019
Supplement Report FDA Received Date05/20/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-