It was reported that this cardiac resynchronization therapy pacemaker (crt-p) system exhibited noise and oversensing on the right ventricular (rv) lead.The patient experienced lightheadedness and palpitations.The noise was recreated with isometrics.In addition, the rv lead impedances were stable around 500 ohms; however, occasionally they were low and out of range measuring less than 200 ohms.Boston scientific technical services (ts) discussed this could be the early stages of a lead insulation breach.Subsequently, a revision procedure was performed and the rv lead was surgically abandoned and replaced.The crt-p remains in service.No additional adverse patient effects were reported.
|