Model Number CQF7598
Medical Device Problem Codes Material Frayed (1262); Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/21/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.(expiry date: 05/2021).
 
Event or Problem Description
It was reported that during an angioplasty procedure in the cephalic arch, the pta balloon allegedly ruptured.Another device was used to complete the procedure.There was no reported patient injury.
 
Event or Problem Description
It was reported that during an angioplasty procedure in the cephalic arch, the pta balloon allegedly ruptured.Another device was used to complete the procedure.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Investigation summary: the device was received for evaluation.The device was inflated, and water was seen leaking from the balloon.Fiber disturbance was noted throughout the balloon.The fibers were stripped, and a compound rupture was noted on the balloon.Therefore, the investigation is confirmed for a compound rupture and fiber disturbance.The definitive root cause for the compound rupture could not be determined based upon the available information.It is unknown if patient and/or procedural issues contributed to the reported event.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.H10: d4 (expiry date: 05/2021); g4 h11 h3; h6 (results, conclusion).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONQUEST 40 PTA DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key8251013
Report Number2020394-2018-02323
Device Sequence Number1558366
Product Code LIT
UDI-Device Identifier00801741060618
UDI-Public(01)00801741060618
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K120660
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCQF7598
Device Catalogue NumberCQF7598
Device Lot NumberRECS2707
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2018
Initial Date Received by Manufacturer 12/19/2018
Supplement Date Received by Manufacturer02/22/2019
Initial Report FDA Received Date01/15/2019
Supplement Report FDA Received Date03/16/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age66 YR
-
-