(b)(4).Fse followed up with the customer over the phone to address the reported event.The customer stated that the issue had been resolved by changing the column and buffers, and that both qc and calibration had passed, and that the patient samples were also normal.The customer emailed the results for review.No further issues were noted.No further action was resolved by field service.A 13 month complaint history review and service history review for similar complaints was performed for the serial number (b)(4) from 17novsep2017 through aware date 17dec2018.There were no similar complaints identified during the search period.The st hba1c aia-pack package insert states the following: quality control: in order to monitor and evaluate the precision of the analytical performance, it is recommended that commercially available control samples be assayed daily.The minimum recommendations for the frequency of running internal control material are: after calibration, three levels of controls are run in order to accept the calibration curve.The three levels of controls are also repeated after calibration when certain service procedures are performed (e.G.Temperature adjustment, sampling mechanism changes, maintenance of the wash probe or detector lamp adjustment or change).After daily maintenance, at least two levels of the control should be run in order to verify the overall performance of the tosoh aia system analyzers.If one or more control sample value(s) is out of the acceptable range, it will be necessary to investigate the validity of the calibration curve and the assay results before reporting patient values.Standard laboratory procedures should be followed in accordance with the regulatory agency under which the laboratory operates.Limitations of the procedure: for diagnostic purposes, the results obtained from this assay should be used in conjunction with other data (e.G., symptoms, results of other tests, clinical impressions, therapy, etc.).The g8 operator's manual under chapter 1- introduction & applications & chapter 6, troubleshooting states the following: storage and stability: the unopened tskgel g8 variant hsi column should be stored at 4-15°c in a cool location away from direct sunlight.The column is stable until the expiration date printed on the label.Columns are warranted for 2500 injections.The most probable cause of the reported event was due to failure of the column.
|
A customer reported that results suddenly dropped during the run with their g8 analyzer.The customer reran the samples on their other g8 analyzer and the results were normal.Next, the customer reran quality control (qc).Level 1 qc results were out of range low and level 3 results were in range.The column count was 422 injections and the filter count was 186.The customer also checked the reagents total area, and pressure; all were within normal ranges.A field service engineer (fse) was dispatched to address the reported event.There was no indication of any patient intervention or adverse health consequences due to the reported event.
|