|
Product complaint # (b)(4).The device history records were reviewed and the manufacturing criteria was met prior to the release of this lot.Evaluation: a labeled winding former, a dispensed needle/suture were returned for analysis.During the visual inspection of sample, the swage and attachment area were noted to be as expected.There were slight marks on the body of the needle which appeared to be by the use of a surgical instrument.The suture was examined and body fluids on barbs were observed; damage near end was noted.However, the two sides of the fixation tab were broken.The manufacturing records were reviewed and the manufacturing/ packaging criteria were met prior to the release of this batch.Per the sample condition of the received needle-suture, the assignable cause of the fixation feature breakage is suggested as an improper handling of the sample.
|