• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 5.0MM TI LOCKING SCR SLF-TPNG W/T25 STARDRIVE 46MM-STERILE; APPLIANCE,FIXATION,NAIL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 5.0MM TI LOCKING SCR SLF-TPNG W/T25 STARDRIVE 46MM-STERILE; APPLIANCE,FIXATION,NAIL Back to Search Results
Model Number 412.217S
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 01/01/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Complainant part is not expected to be returned for manufacturer review/investigation.(b)(4) the reported event required medical/surgical intervention to preclude permanent damage to a body structure.The investigation could not be completed; no conclusion could be drawn, as no product was received.A review of the device history records has been requested.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
On (b)(6) 2018 the pt had a femoral neck system implanted for a femoral neck fracture.The pt subsequently had multiple falls post-operative which resulted in his femoral head rotating and the implant followed.The revision surgery was performed on (b)(6) 2018 and converted to a total hip.This report is for (1) 5.0mm ti locking scr slf-tpng w/t25 stardrive 46mm-sterile.This report is 3 of 4 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2018 patient had a femoral neck system implanted for a femoral neck fracture.The patient subsequently had multiple falls post-operative which resulted in his femoral head rotating and the implant followed.The revision surgery was performed on (b)(6) 2018 and converted to a total hip.Patient and surgical outcome were not reported.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device history records review was completed for part: 412.217s, lot: l790188.Manufacturing location: mezzovico, release to warehouse date: mar 01, 2018, expiry date: feb 01, 2028.The device history record shows this lot was processed through the normal manufacturing and inspection operations with no rework or nonconformities noted.This lot met all dimensional and visual criteria at the time of release with no issues documented during the manufacturing process.Review of the device history records showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.0MM TI LOCKING SCR SLF-TPNG W/T25 STARDRIVE 46MM-STERILE
Common Device Name
APPLIANCE,FIXATION,NAIL
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key8253304
Report Number2939274-2019-55886
Device Sequence Number16691674
Product Code KTT
UDI-Device Identifier10886982275448
UDI-Public(01)10886982275448
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K000682
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number412.217S
Device Catalogue Number412.217S
Device Lot NumberL790188
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/21/2018
Supplement Date Received by Manufacturer01/22/2019
02/18/2019
Initial Report FDA Received Date01/16/2019
Supplement Report FDA Received Date02/04/2019
02/19/2019
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age83 YR
Patient Weight79
-
-