| Catalog Number |
0066204 |
| Medical Device Problem Code |
False Negative Result (1225)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/24/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
On (b)(6) 2018 immucor technical support used a remote electronic connection method to review results for the issue described.The camera report appeared acceptable and no instrument problem was identified.On (b)(6) 2018, immucor tested selected cells from retention capture-r ready-id lot id369 in manual capture with anti-k lot k012512f.Controls performed as expected and cells 6 and 8 (k positive) resulted positive as expected.Retention product performed as expected.No sample or product was returned to immucor for evaluation.The immucor internal report record number for this report is (b)(4).
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018 a customer reported that they received unexpected negative reactions with capture-r ready-id on an echo v2.0 instrument.
|
| |
|