Model Number 497-28-700
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 12/27/2018
Type of Reportable Event Serious Injury
Event or Problem Description
Second revision surgery - due to dislocation.
 
Additional Manufacturer Narrative
The reason for this second revision surgery was dislocaiton.The previous surgery and the revision detailed in this investigation occurred over 12 days apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.Initial or prolonged hospitalization was required.This investigation is limited in scope as limited information was provided to djo surgical - austin for examination and the part associated with this investigation was not returned to djo surgical for review.A review of the implant device history records (dhr) shows that the reported components used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports associated with the product that may have contributed to the event.The device and its applicable concomitant devices were within their respective expiration dates at the time of use during the previous surgery.Customer complaint history of the reported device (s) showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to dislocation.There were no findings during this investigation that indicate that the reported device was the source or had a direct connection with the event.There are multiple factors that may contribute to the event that are outside the control of djo surgical are inadequate soft tissue, patient bone deterioration, excessive range of motion, patient activities or trauma.Due to the short time between previous and revision surgery, it is also possible that the event may have been occurred due to improper implant selection or patient non-compliance with medical instructions.No additional information was submitted with the complaint regarding pre-existing conditions of the patient or any activities that may have contributed to the event and hence a definitive root cause cannot be determined.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FMP HIP
Common Device Name
HEAD, FEMORAL, COCR 28/+7MM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key8254774
Report Number1644408-2019-00039
Device Sequence Number13931357
Product Code LPH
UDI-Device Identifier00888912087292
UDI-Public(01)00888912087292
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K932481
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/24/2022
Device Model Number497-28-700
Device Catalogue Number497-28-700
Device Lot Number635C1141
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/27/2018
Supplement Date Received by Manufacturer01/25/2019
Initial Report FDA Received Date01/16/2019
Supplement Report FDA Received Date01/29/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
412-02-047, LOT 993B1091
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age76 YR
-
-