• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. HS FOCUS 9CM CURVED SHEAR; INSTRUMENT, ULTRASONIC SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. HS FOCUS 9CM CURVED SHEAR; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Catalog Number FCS9
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch # m90a3a.Investigation summary: the device was returned with the distal tip of the blade broken off and not returned.The remaining blade portion was scratched, and with evidence of contact with metal in or out of the operative field.This blade tip portion may have broken off of the device during transport to our analysis site.During functional testing on gen11 an alert screen was displayed.A probable cause of the device stop activating and display an alert screen is blade damage.Probable causes of blade damage, including breakage, are external contact during pre-op or general use, blade contact with other devices, staples or clips during the procedure or using any means other than the blade wrench to attach or detach the blade.Once minor blade damage has occurred, subsequent activations may increase damage severity and can result in not passing the pre-run test followed by an alert screen, such as ¿tighten assembly¿, ¿blade error detected¿ or "relaxed pressure on blade" followed by a ¿replace instrument¿ screen later in the procedure, and continued usage can result in a broken blade.A batch history record review was performed and a protocol was created during the manufacturing of this batch but was not related to the event description.Additionally, no defects or nc related to the complaint were found during the manufacturing process.
 
Event or Problem Description
It was reported that during an unknown procedure, the generator alarmed that blade pressure need to be low.No patient consequence reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HS FOCUS 9CM CURVED SHEAR
Common Device Name
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key8255217
Report Number3005075853-2019-15999
Device Sequence Number19404564
Product Code LFL
UDI-Device Identifier10705036001959
UDI-Public10705036001959
Combination Product (Y/N)N
PMA/510(K) Number
K063192
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2020
Device Catalogue NumberFCS9
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/02/2019
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/03/2019
Initial Report FDA Received Date01/16/2019
Patient Sequence Number1
-
-