• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD; INC. PORTEX BIVONA FLEXTEND¿ TTS¿ NEONATAL AND PEDIATRIC TRACHEOSTOMY TUBES; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD; INC. PORTEX BIVONA FLEXTEND¿ TTS¿ NEONATAL AND PEDIATRIC TRACHEOSTOMY TUBES; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Catalog Number 67PFP60
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 12/13/2018
Type of Reportable Event Serious Injury
Event or Problem Description
Information was received indicating that a patient with a smiths medical portex bivona flextend¿ tts¿ neonatal and pediatric tracheostomy tube was reported to be experiencing secretions and air coming from the stoma when the cuff is not fully inflated.Subsequently, the trach tube was removed and the nurse filled the trach with water.A pinhole tear in the cuff was observed.There were no reported adverse patient effects.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX BIVONA FLEXTEND¿ TTS¿ NEONATAL AND PEDIATRIC TRACHEOSTOMY TUBES
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL ASD; INC.
6000 nathan lane n
minneapolis, MN 55442
MDR Report Key8259005
Report Number3012307300-2019-00232
Device Sequence Number19359393
Product Code JOH
UDI-Device Identifier10351688518538
UDI-Public10351688518538
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K923878
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 01/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number67PFP60
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/19/2018
Initial Report FDA Received Date01/17/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-