• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC MAXFRAME(TM) FULL RING 150MM ALUMINUM; APPLIANCE,FIXATION,NAIL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC MAXFRAME(TM) FULL RING 150MM ALUMINUM; APPLIANCE,FIXATION,NAIL Back to Search Results
Model Number 03.312.150
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.A review of the device history records has been requested.(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on an unknown date, the maxframe full ring broke on the patients frame which was being used to treat a distal tibia fracture.The original implant procedure was on (b)(6) 2018.The broken ring was replaced on (b)(6) 2019, with a 150mm 5/8 ring and bridge plate.The patient also broke two (2) smooth wires and one (1) olive wires.The wires were also replaced with new ones.It is unknown it there was surgical delay.The procedure was successfully completed.Patient status is unknown.This report is for one (1) maxframe(tm) full ring 150mm aluminum.This is report 1 of 4 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device history part number 03.312.150, lot number: h420922, date of manufacture: 09-19-17, place of manufacture: brandywine plant.Part expiration date: n/a (non-sterile).Device history batch null, device history review review of the device history record(s) showed that there were no issues during the manufacture of this product, which would contribute to this complaint condition.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAXFRAME(TM) FULL RING 150MM ALUMINUM
Common Device Name
APPLIANCE,FIXATION,NAIL
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key8261151
Report Number2939274-2019-55924
Device Sequence Number183720
Product Code KTT
UDI-Device Identifier10886982260819
UDI-Public(01)10886982260819
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
PMA/510(K) Number
K161417
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number03.312.150
Device Catalogue Number03.312.150
Device Lot NumberH420922
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/27/2018
Supplement Date Received by Manufacturer02/05/2019
02/14/2019
Initial Report FDA Received Date01/18/2019
Supplement Report FDA Received Date02/12/2019
02/15/2019
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-