Model Number VPR-GW-FT14
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code Perforation of Vessels (2135)
Date of Event 01/04/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the reported device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.The material inspection report for this guide wire lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.(b)(4).
 
Event or Problem Description
During a procedure with a diamondback peripheral orbital atherectomy device (oad), a perforation of the left anterior tibial artery occurred.Lesions in the superficial femoral artery and popliteal artery were successfully treated with orbital atherectomy and balloon angioplasty.At the end of the procedure, a perforation was noted in the left anterior tibial artery, and a blood pressure cuff was applied to the area.The patient was doing well following the procedure.The physician attributed the perforation to user error when advancing the balloon over the guide wire resulting in the guide wire advancing into a small side branch and not a device malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM
Common Device Name
PERIPHERAL ATHERECTOMY DEVICE
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old highway 8 nw
saint paul MN 55112
MDR Report Key8261481
Report Number3004742232-2019-00028
Device Sequence Number16184965
Product Code MCW
UDI-Device Identifier10852528005312
UDI-Public(01)10852528005312(17)200731(10)236515
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K133399
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2020
Device Model NumberVPR-GW-FT14
Device Catalogue NumberVPR-GW-FT14
Device Lot Number236515
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/04/2019
Initial Report FDA Received Date01/18/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-