| Model Number |
MT20649-BLU |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/02/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
Dexcom as made aware on (b)(4) 2019, that on (b)(6) 2019, an err hw rf, err 67, and err 189 was displayed on the receiver.No product or data was not available for evaluation.Confirmation of the error icon and a probable cause could not be determined.No additional event or patient information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
The receiver was returned for evaluation.An external visual inspection was performed and passed.A receiver charge and boot test was performed and passed.The receiver data log was downloaded for review and hardware and firmware errors along with ¿screen error alarm¿ were observed.The receiver case was opened for an internal visual inspection and it passed.The reported event of an error icon display was confirmed during log review.A probable cause could not be determined.
|
| |
|