| Model Number |
FGS-0400 |
| Medical Device Problem Codes |
Entrapment of Device (1212); Poor Quality Image (1408)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, the capsule was stuck in the stomach and the study had poor picture quality.There was no reported patient outcome.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: evaluation summary: this report is based on information provided by medtronic investigation personnel and the sample that arrived.One bravo capsule and one bravo delivery device were received for evaluation.A review of the product expiration date discovered this product was used before the expiration date.The returned sample met specification as received by medtronic.The visual inspection found no notable conditions.The customer reported bravo fail to attach to patient's esophagus.The investigation of the returned equipment did not identify anything that would have caused or contributed to the reported event.The investigation found the device to function normally and within specifications.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, the capsule was stuck in the stomach and the study had poor picture quality.There was no reported patient outcome.
|
| |
|