If additional information should become available, a supplemental medwatch will be submitted accordingly.Udi: (b)(4).The complaint device is not being returned, therefore unavailable for a physical evaluation.A review into the depuy synthes mitek complaints system revealed no other complaints for this lot of devices that were released to distribution.We cannot discern a root cause for the reported failure mode.At this point in time, no corrective action is required and no further action is warranted.This file will remain receptive to any potential forthcoming information received that is pertinent and germane to this issue.A non-conformance search was performed for this product code 254681, lot #155075r combination and no non-conformances were identified.No further information regarding the technique or instruments used has been provided to determine a root cause for this failure.If any additional information is obtained, this complaint will be re-opened to capture that information.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.This report is being filed from the etq complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
|
|
It was reported by the sales rep that during an anterior cruciate ligament surgical procedure, it was observed that the screw in the customer's femoral intrafix sheath & screw system 7.5mm cracked and fractured during insertion parallel to the screw.The sales rep stated that the surgeon removed the screw with no further issues and that no debris was left in the patient.The sales rep reported that the surgeon completed the procedure with a smaller size 6 hard bone screw using the same sheath and bone hole with no patient consequences or delays.The sales rep stated that the patient's bone quality was hard bone.The sales rep stated that the device was discarded by the facility.There was patient involvement reported.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|