SMITH & NEPHEW, INC. GII PATELLAR REAMER SHAFT; PROST, KNEE, PATEL/FEMTIBIAL, SEMI-CONS, POROUS, COAT, POLY/METAL/POLY
|
Back to Search Results |
|
| Catalog Number |
71440324 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/27/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during surgery they were reaming the femur box for a ps femur and when they got done reaming we noticed the pointed end of the reamer shaft had broken off.No delay was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The drill point disassembled from the device and was not returned.The device was manufactured in 2013 and exhibits signs of extensive wear/usage.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|