SMITH & NEPHEW, INC. GII PATELLAR REAMER SHAFT; PROST, KNEE, PATEL/FEMTIBIAL, SEMI-CONS, POROUS, COAT, POLY/METAL/POLY Back to Search Results
Catalog Number 71440324
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/27/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during surgery they were reaming the femur box for a ps femur and when they got done reaming we noticed the pointed end of the reamer shaft had broken off.No delay was reported.
 
Additional Manufacturer Narrative
The associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The drill point disassembled from the device and was not returned.The device was manufactured in 2013 and exhibits signs of extensive wear/usage.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII PATELLAR REAMER SHAFT
Common Device Name
PROST, KNEE, PATEL/FEMTIBIAL, SEMI-CONS, POROUS, COAT, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8265275
Report Number1020279-2019-00222
Device Sequence Number16733672
Product Code MBH
UDI-Device Identifier03596010214997
UDI-Public03596010214997
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K121393
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number71440324
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/27/2018
Supplement Date Received by Manufacturer12/27/2018
Initial Report FDA Received Date01/21/2019
Supplement Report FDA Received Date02/15/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-