Catalog Number 5546-A-302
Device Problem Loosening of Implant Not Related to Bone-Ingrowth (4002)
Patient Problems Injury (2348); Inadequate Osseointegration (2646)
Event Date 12/26/2018
Event Type  Injury  
Manufacturer Narrative
An event regarding tibial loosening involving a triathlon augment was reported.The event was not confirmed.Method & results: product evaluation and results: not performed as no product was returned for evaluation.Clinician review: not performed as medical records were not provided.Product history review: indicated all devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Additional information including operative reports, progress notes, x-rays and return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
 
Event Description
As reported: "patient's left tibial component was revised due to loosening¿all tibial components [12x50 stem, size 3 universal baseplate, two augments, 3x16 ts insert¿were explanted and revised to triathlon ts.Tibial side as a whole was loosened, no specific implant(s) cited".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRI RM/LL TIB AUG SZ3 10MM
Type of Device
KNEE JOINT PATELLOFEMOROTIBIAL METAL/POLYMER POROUS-COATED UNCEMENTED PROSTHESIS
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
keyla colon
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key8265349
MDR Text Key133706694
Report Number0002249697-2019-00239
Device Sequence Number1
Product Code MBH
UDI-Device Identifier07613327026474
UDI-Public07613327026474
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Physician
Type of Report Initial
Report Date 01/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/21/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date06/30/2013
Device Catalogue Number5546-A-302
Device Lot NumberN3K43A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/26/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/16/2008
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age56 YR
-
-