| Model Number |
ZA9003 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
12/19/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If implanted, give date: not applicable, as lens was not implanted.If explanted, give date: not applicable, as lens was not implanted.All pertinent information available to johnson and johnson surgical vision has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the lens had broken haptic which was observed when removing from packaging.There was no patient involvement.No further information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the package returned was received in the manufacturing site.However, the contents in the package did not include the iols (intraocular lens) for this complaint folder.The product testing could not be performed as the product was not returned.Therefore, a product evaluation is not possible.The reported issue could not be verified.Manufacturing record review: the manufacturing records for the product was reviewed.The product was manufactured and released according to specifications.A search revealed that no additional investigation has been received for this production order.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-010215.
|
| |
|