| Model Number |
MMT-722NAS |
| Medical Device Problem Code |
Power Problem (3010)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/15/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.The insulin pump involved in this event is the 640g insulin infusion pump, which is not marketed in the united states.However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.
|
| |
|
Event or Problem Description
|
|
It was reported that the insulin pump had off no power without any alert.The customer¿s blood glucose level was unknown at the time of the incident.Troubleshooting was not performed for power loss alarm.The insulin pump will not be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with blank display due to corroded battery tube.Unable to perform operating currents, self-test, a21 error test and displacement test or verify low battery, battery out limit alarms due to blank display.Device had broken battery tube threads.
|
| |
|