• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US DURALOC SECTOR SERIES 48OD; DURALOC IMPLANT : HIP METAL ACETABULAR CUPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US DURALOC SECTOR SERIES 48OD; DURALOC IMPLANT : HIP METAL ACETABULAR CUPS Back to Search Results
Catalog Number 124580055
Medical Device Problem Code Fitting Problem (2183)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/22/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Duraloc 48mm acetabular did not give the press fit needed for fixing.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Investigation summary: examination of the returned acetabular cup finds nothing outward to suggest product error.A worldwide complaint database search found no additional related reports against the provided product code/lot code combination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURALOC SECTOR SERIES 48OD
Common Device Name
DURALOC IMPLANT : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key8266683
Report Number1818910-2019-81713
Device Sequence Number2370819
Product Code LPH
UDI-Device Identifier10603295019640
UDI-Public10603295019640
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K961186
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number124580055
Device Lot Number8057891
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/25/2019
Initial Date Received by Manufacturer 12/28/2018
Supplement Date Received by Manufacturer03/13/2019
Initial Report FDA Received Date01/22/2019
Supplement Report FDA Received Date03/14/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
-
-