| Model Number |
1203012 |
| Medical Device Problem Code |
Particulates (1451)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/01/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned.It has not yet been received.
|
| |
|
Event or Problem Description
|
|
The event involved a plum set on an unspecified date that was reported to have impurities inside the tube.There was no report of human harm, nor adverse event, nor medical intervention.
|
| |
|
Additional Manufacturer Narrative
|
|
One (1) used list number 120300412 lifeshield primary plumset clave port clave y-site was received for evaluation.As received, particulate is observed within the tubing of the returned used plum set.The particulate was isolated from the tubing and submitted for ft-ir analysis.The ft-ir revealed that the particulate closely matched a dextrose solution.The infusate administered was not provided by the customer.The probable cause of the observed particulate is infusate residual accumulated during use.A batch record review was performed to lot# 810465h, list# 120300412 and no discrepancies that may have contributed to a complaint of this nature were found.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected information in date received by manufacturer.
|
| |
|