| Model Number |
53312 |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Codes |
Hemorrhage/Blood Loss/Bleeding (1888); Patient Problem/Medical Problem (2688)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the diameter of the magic 3 catheters vary in size.Complainant allegedly experienced self-limited bleeding with the use of some of the catheters due to the varying diameters.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was unconfirmed.Visual evaluation noted 3 unopened samples in the original packaging.No damage was noted on the exterior of the packaging.The dip line and hydrophilic coating were both present.All catheters tested measured to be 12fr.The lot number is unknown therefore the device history record could not be reviewed.The instructions for use were found adequate and state the following: ¿for urological use only.Urinary catheters are intended for use for bladder management including urine drainage, collection and measurement.The devices are passed to the urinary bladder via the urethra.Warnings: this is a single use device.Do not reuse.Reuse of a single use device increases the risk of catheter acquired urinary tract infections.How to prepare and use your urinary catheter 1.Wash your hands thoroughly with soap and water.2.Peel open the pack at the funnel end to expose approximately 2¿- 4¿ of the catheter.Don¿t remove the catheter yet.3.Wash the area around the meatus before catheterizing.4.Wash your hands again.5.Using the catheter: with sure-grip¿ sleeve: a.Hold the sure-grip¿ sleeve with your dominant hand and squeeze it to hold the catheter shaft as you remove the catheter from the pack.B.Next, hold the catheter funnel above the sure-grip¿ sleeve with your other hand and slide the sure-grip¿ sleeve down the shaft, stopping at about 6¿ from the tip.Release the funnel.C.Using the sure-grip¿ sleeve to hold the catheter firmly, lubricate catheter tip with lubricating gel.D.While gripping the catheter with the sure-grip¿ sleeve, gently pass the tip of the catheter into your urethra until the sure-grip¿ sleeve nears the meatus.If there is no urine flow, loosen the tension on the sure-grip¿ sleeve to allow it to slide back towards the funnel end.Re-grip sure-grip¿ sleeve and continue to insert the catheter into the urethra.Repeat until urine starts to flow.Without sure-grip¿ sleeve: a.Remove the catheter from the pack.B.Lubricate the catheter tip with lubricating gel.C.Gently pass the tip of the catheter into your urethra and advance the catheter until urine starts to flow.6.Try to keep the catheter steady until urine stops flowing.When urine stops flowing, slowly withdraw the catheter, stopping if flow starts again, until the last few drops have drained.7.Finish by disposing of the catheter and its packaging.8.Wash your hands with soap and water.".
|
| |
|
Event or Problem Description
|
|
It was reported that the diameter of the magic 3 catheters vary in size.Complainant allegedly experienced self-limited bleeding with the use of some of the catheters due to the varying diameters.
|
| |
|