• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SPINE INC XPDM THORACIC PEDICLE PRB, CRV; PROBE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SPINE INC XPDM THORACIC PEDICLE PRB, CRV; PROBE Back to Search Results
Model Number 279702040
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Excessive Tear Production (2235); Spinal Cord Injury (2432)
Date of Event 12/31/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).(b)(6).(b)(4) device is available for evaluation.Investigation will be conducted.Follow up will be filed with the investigation results.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that, during surgery, dr.(b)(6) was using the expedium pedicle probe to prepare the pedicle for screw insertion into the left side of s1 (sacral 1).He had just commented how hard the patient's bone was during insertion of other screws.At this time the distal end of the probe snapped off.He was able to easily recover broken fragment, but a small dural tear was created.He had to complete a dural repair with suture and fibrin sealant.Part number: 297-02-040, lot #: nw140086.Was it delayed? yes.How long? 20min.Action taken - dural repair - replacement of broken instrument.Was procedure completed? yes.Fragments: yes.Easily removed: yes.Patient "consequenses": unknown.Other medical intervention: dural repair.
 
Additional Manufacturer Narrative
(b)(4).Device was not returned for evaluation.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.All complaint trends will be evaluated as a part of the depuy spine monthly complaint review meeting.Without the return of the device, we are unable to confirm the reported issue or identify the root cause.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer and no systemic trends were found, this complaint file will be closed with no further action required.Should more information and/or the sample be provided at a later time, this complaint will be reopened and device evaluated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XPDM THORACIC PEDICLE PRB, CRV
Common Device Name
PROBE
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
MDR Report Key8267941
Report Number1526439-2019-51263
Device Sequence Number2202718
Product Code HXB
UDI-Device Identifier10705034198446
UDI-Public(01)10705034198446
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number279702040
Device Catalogue Number279702040
Device Lot NumberNW140086
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/31/2018
Supplement Date Received by Manufacturer02/26/2019
Initial Report FDA Received Date01/22/2019
Supplement Report FDA Received Date02/26/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
Patient Weight115
-
-