• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US 4590 FMS SOLO IRRIGATION PUMP -NS; DISTENSION UNIT, FLUID, ARTHROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US 4590 FMS SOLO IRRIGATION PUMP -NS; DISTENSION UNIT, FLUID, ARTHROSCOPIC Back to Search Results
Catalog Number 284590
Device Problems Increase in Pressure (1491); Pressure Problem (3012)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/03/2017
Event Type  malfunction  
Manufacturer Narrative
If additional information should become available, a supplemental medwatch will be submitted accordingly.(b)(4).The device was received at the service center and repaired.The unit was evaluated and the reason for return (incorrect pressure) was verified.Replaced worn finger on complete pressure adjuster (preventive maintenance) with tip replacement kit to address (incorrect pressure.) the unit passed all diagnostic tests, functional tests, and is fully operational.At this time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family to monitor the extent to which this complaint is observed in the field.This report is being filed from the (b)(4) as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
 
Event Description
It was reported by the customer that during a shoulder arthroscopy surgical procedure, it was observed that the 4590 fms solo irrigation pump device was displaying incorrect pressure.According to the reporter, the pressure was either too high or too low.There was a delay in the surgical procedure of less than 15 minutes.An identical spare device was used to complete the procedure.There was patient involvement reported.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4590 FMS SOLO IRRIGATION PUMP -NS
Type of Device
DISTENSION UNIT, FLUID, ARTHROSCOPIC
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6013142063
MDR Report Key8269021
MDR Text Key134457654
Report Number1221934-2019-56089
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705020614
UDI-Public10886705020614
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K002040
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 01/03/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/22/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number284590
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/13/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/03/2017
Date Device Manufactured06/13/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-