ENCORE MEDICAL L.P. ALTIVATE ANATOMIC SHOULDER; ALTIVATE ANATOMIC, OFFSET HUMERAL HEAD, 50X18 Back to Search Results
Catalog Number 520-50-318
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/07/2019
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the head implant disassociating from the humeral neck.The surgeon removed and replaced the original head and neck.
 
Additional Manufacturer Narrative
(b)(4).Manufacturer narrative: the reason for this revision surgery was implant disassociation.The previous surgery and the surgery detailed in this investigation occurred 5.3 months apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.This investigation is limited in scope as the item(s) associated with this investigation was not returned to djo surgical - austin for examination.A review of the implant device history records (dhr), shows that the reported component(s) used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports associated with the product(s) that may have contributed to the reported event.The device(s) was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device(s) showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to implant disassociation.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.There were no findings during this investigation that indicate that the reported device(s) was defective.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTIVATE ANATOMIC SHOULDER
Common Device Name
ALTIVATE ANATOMIC, OFFSET HUMERAL HEAD, 50X18
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key8270878
Report Number1644408-2019-00053
Device Sequence Number1952606
Product Code KWS
UDI-Device Identifier00190446176299
UDI-Public(01)00190446176299
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K162024
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/06/2024
Device Catalogue Number520-50-318
Device Lot Number958U1040
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/07/2019
Supplement Date Received by Manufacturer02/07/2019
Initial Report FDA Received Date01/23/2019
Supplement Report FDA Received Date02/22/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
520-50-318, LOT 958U1040
Outcome Attributed to Adverse Event Required Intervention;
Patient Age50 YR
-
-