The device history record was reviewed.The original lot number of the plate was 1610300030.There were no anomalies.From the x-rays, the plate broke where there would be the highest stress.The patient displayed a malunion.The bones were not fused.The plate broke after a year of withstanding in-vivo forces.Upon review of the x-rays from a second doctor, dr.(b)(6) said: "it's too plantarflexed.The distal screw is barely biting anything and the d to p screw is too close to the joint", which created a situation where fusion would be unlikely.Therefore the root cause of the plate breakage was due to surgeon technique.A report was conducted regarding the blackness around the plate, the report determined that it was "not" "metalosis" nor infection, it was just black pigment.(b)(4).
|
Mtp plate(mfp 012, lot 1701320000) implanted on (b)(6) 2018 broke; revision surgery happened on (b)(6) 2019.During the removal, there was also a report of blackness surrounding the plate, which the doctor believed could be "metalosis".
|