Model Number 3716 |
Device Problem
Temperature Problem (3022)
|
Patient Problems
Burning Sensation (2146); Discomfort (2330)
|
Event Type
Injury
|
Manufacturer Narrative
|
Concomitant medical products and therapy dates: model: 3701, scs charging system, therapy date: not applicable.The investigation for capa (b)(4) associated with heating while charging (pocket heating) concluded that the charger was capable of transferring energy to the ipg at a rate that would cause heating of the ipg and/or charging wand of sufficient elevated temperature to cause pain and burns.The heating while charging was determined to be exacerbated by off-axis charging of shallow implanted ipgs and that all chargers were capable of elevated heating.
|
|
Event Description
|
Device 1 of 2: reference mfr.Report#: 1627487-2019-01124.It was reported the patient had been experiencing uncomfortable heating at their pocket site while recharging their ipg.The patient was able to address the issue with a replacement/different model charging system.
|
|
Event Description
|
Device 1 of 2; reference mfr.Report#: 1627487-2019-01124.Follow-up revealed the patient's ipg was explanted and replaced due to the pocket heating issue that was initially reported, along with the ipg allegedly reaching end-of-life.Surgical intervention addressed the patient's issue.
|
|