ST. JUDE MEDICAL - NEUROMODULATION EON RECHARGEABLE IPG, 16-CHANNEL; SCS IPG Back to Search Results
Model Number 3716
Device Problem Temperature Problem (3022)
Patient Problems Burning Sensation (2146); Discomfort (2330)
Event Type  Injury  
Manufacturer Narrative
Concomitant medical products and therapy dates: model: 3701, scs charging system, therapy date: not applicable.The investigation for capa (b)(4) associated with heating while charging (pocket heating) concluded that the charger was capable of transferring energy to the ipg at a rate that would cause heating of the ipg and/or charging wand of sufficient elevated temperature to cause pain and burns.The heating while charging was determined to be exacerbated by off-axis charging of shallow implanted ipgs and that all chargers were capable of elevated heating.
 
Event Description
Device 1 of 2: reference mfr.Report#: 1627487-2019-01124.It was reported the patient had been experiencing uncomfortable heating at their pocket site while recharging their ipg.The patient was able to address the issue with a replacement/different model charging system.
 
Event Description
Device 1 of 2; reference mfr.Report#: 1627487-2019-01124.Follow-up revealed the patient's ipg was explanted and replaced due to the pocket heating issue that was initially reported, along with the ipg allegedly reaching end-of-life.Surgical intervention addressed the patient's issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EON RECHARGEABLE IPG, 16-CHANNEL
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer Contact
andrea deitz
6901 preston road
plano, TX 75024
9723098000
MDR Report Key8272012
MDR Text Key133901819
Report Number1627487-2019-01123
Device Sequence Number1
Product Code LGW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date 03/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/23/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/01/2008
Device Model Number3716
Device Catalogue Number3716
Device Lot Number50958
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/15/2019
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/18/2006
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487-07262012-001-C
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Weight68
-
-