| Model Number |
MMT-523RNAS |
| Medical Device Problem Code |
Infusion or Flow Problem (2964)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/03/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Customer reported via phone call the insulin pump had priming anomaly.Insulin pump was squirting and does not stop rewinding.Troubleshooting was not completed.Customer is not using the insulin pump.The insulin pump will be returning for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed displacement test.However unit alarmed motor error during basic occlusion test and unable to prime during prime test due to moisture damage force sensor resistance.Unable to perform occlusion test or excessive no delivery test due to motor error alarms.The motor was tested outside the device and passed.
|
| |
|