| Model Number |
MMT-723NAP |
| Medical Device Problem Codes |
Failure to Cycle (1142); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
01/21/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that they experienced high blood glucose and insulin pump alarmed compromised force sensor system.The customer¿s blood glucose level was 528 mg/dl at the time of the incident.The customer did not provide details regarding symptoms and treatment of high blood glucose.The customer also reported that the insulin was squirting during manual prime process.The customer stated that the drive support cap was normal.The customer was advised that the insulin pump needed to be replaced and was advised to discontinue use of the insulin pump and to revert to the backup plan.The insulin pump will be returned for analysis.
|
| |
|
Search Alerts/Recalls
|