| Model Number |
6533 |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/10/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: returned product consisted of a nc quantum apex balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed multiple minor kinks along the hypotube.Microscopic examination revealed that the guidewire lumen is buckled 47mm from the tip and is approximately 2mm long.There is a hole in the guidewire lumen in the buckled section approximately 48mm from the tip.The balloon is still tightly folded.The guide wire used in the clinical procedure was not returned with the device so a test guide wire was used for functional testing.A.014 in test guide wire was advanced into the guide wire lumen and was unable to pass the buckled section.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 04-jan-2019.It was reported that loading difficulties were encountered.The target lesion was located in a coronary artery.A 20mm x 3.50mm nc quantum apex and a 2.50mm x 12mm emerge balloon catheters were selected for use; however, both devices could not be loaded over the wire.The procedure was completed with a non-bsc device.No patient complications were reported since the devices did not enter the patient's body.However, returned device analysis revealed hole in shaft material.
|
| |
|