• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US SUPER QA+ #2 ETH CP-2 *EA; SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US SUPER QA+ #2 ETH CP-2 *EA; SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE Back to Search Results
Catalog Number 212032
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/27/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If additional information should become available, a supplemental medwatch will be submitted accordingly.Udi: (b)(4).The complaint device was received and evaluated.Visual observation confirms that the inserter tip is bent outwards on one side.There was no damage to the anchor itself when inspected.A definite root cause cannot be determined; it is a possibility that excess torque was applied when inserting the anchor at off angle which caused the inserter tip to bend.Other than this possibility, a root cause cannot be determined.Further, a review into the depuy synthes mitek complaints system revealed no similar complaint of any kind for this lot of devices that were released to distribution.At this point in time, no corrective action is required and no further action is warranted.A non-conformance search was performed for this product code 212032-lot #l165772 combination and no non-conformances were identified.No further information regarding the technique or instruments used has been provided to determine a root cause for this failure.If any additional information is obtained, this complaint will be re-opened to capture that information.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.This report is being filed from the etq complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
 
Event or Problem Description
It was reported by the customer that during an unspecified surgical procedure of the leg and ankle repair, it was observed that the super qa+ #2 eth cp-2 *ea device was bent when it was opened.There was no delay in the surgical procedure as an identical spare device was used to complete the procedure.There was patient involvement reported.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key8274124
Report Number1221934-2018-53588
Device Sequence Number11250931
Product Code JDR
UDI-Device Identifier10886705001804
UDI-Public10886705001804
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
K930893
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2021
Device Catalogue Number212032
Device Lot NumberL165772
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/06/2017
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/27/2017
Initial Report FDA Received Date01/24/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-