| Catalog Number |
72203270 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/14/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that during the surgery the ball joint on ahtb attachment would not lick in place adequately when the patients leg was in the boot.The surgical assistant had to hold the leg in the required position for the surgeon.No back up device was available.No patient injury or other complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An evaluation was performed by smith and nephew and could not confirm the customer complaint for the ball joint on ahtb attachment would not click in place adequately.This device was manufactured in 2013.A visual inspection was performed and showed cosmetic damage only.Functional inspection was performed and showed when the hip distractor was attached to the table attachment the device connected properly.No manufacturing related defects were observed.
|
| |
|