Catalog Number 72203270
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/14/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).
 
Event or Problem Description
It was reported that during the surgery the ball joint on ahtb attachment would not lick in place adequately when the patients leg was in the boot.The surgical assistant had to hold the leg in the required position for the surgeon.No back up device was available.No patient injury or other complications were reported.
 
Additional Manufacturer Narrative
An evaluation was performed by smith and nephew and could not confirm the customer complaint for the ball joint on ahtb attachment would not click in place adequately.This device was manufactured in 2013.A visual inspection was performed and showed cosmetic damage only.Functional inspection was performed and showed when the hip distractor was attached to the table attachment the device connected properly.No manufacturing related defects were observed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AHTB UNIVERSAL HIP DISTRACTOR
Common Device Name
APPARATUS, TRACTION, NON-POWERED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
MDR Report Key8275893
Report Number3003604053-2019-00007
Device Sequence Number16525832
Product Code HST
UDI-Device Identifier00885554024821
UDI-Public00885554024821
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number72203270
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/25/2019
Initial Date Received by Manufacturer 01/14/2019
Supplement Date Received by Manufacturer03/19/2019
Initial Report FDA Received Date01/24/2019
Supplement Report FDA Received Date03/20/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-