SMITHS MEDICAL ASD, INC. PORTEX COMBINED SPINAL EPIDURAL CUSTOM TRAY; ANESTHESIA CONDUCTION KIT Back to Search Results
Catalog Number A2744-17/27
Medical Device Problem Codes Difficult to Insert (1316); Material Too Rigid or Stiff (1544)
Health Effect - Clinical Codes Infiltration into Tissue (1931); Perforation (2001)
Date of Event 12/25/2018
Type of Reportable Event Malfunction
Event or Problem Description
Information was received that a smiths medical portex combined spinal epidural custom tray was too stiff, causing vascular penetration.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
One epidural catheter and connector were received for evaluation.Upon visual inspection of the flat filter and connector, no discrepancies were noted.Visual inspection of the returned catheter revealed that the tip was bent approximately 5mm from the distal end.However, the catheter tip appeared to be smooth and rounded.All catheter eyes were noted to be present and were properly formed.A dried substance was present throughout the inside of the catheter.As functional testing, attempts were made to push fluid through the catheter/connector assembly (using a fluid filled syringe) to observe for flow from the catheter eyes.This test was not successful due to the dried substance inside the catheter, impeding the flow.The presence of the dried substance may support the customer's claim for the occurrence of an intravascular penetration.However, the returned catheter was found to be within specification and was fully formed with a smooth, rounded tip.The cause of the reported issue was not determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX COMBINED SPINAL EPIDURAL CUSTOM TRAY
Common Device Name
ANESTHESIA CONDUCTION KIT
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
MDR Report Key8276016
Report Number3012307300-2019-00365
Device Sequence Number11282933
Product Code CAZ
UDI-Device Identifier30351688059301
UDI-Public30351688059301
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/01/2020
Device Catalogue NumberA2744-17/27
Device Lot Number3713218
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/27/2018
Supplement Date Received by Manufacturer02/19/2019
Initial Report FDA Received Date01/24/2019
Supplement Report FDA Received Date03/19/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-