| Catalog Number |
A14BX040150170 |
| Medical Device Problem Codes |
Burst Container or Vessel (1074); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Patient Problem/Medical Problem (2688)
|
| Date of Event |
01/16/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The physician was intending to use a rapidcross to treat a soft tissue lesion in the upper extremity.The device was prepped with no issues.No embolic protection was used.During inflation with a boston scientific inflation device, a balloon burst occurred at nominal atm.The balloon detached from the shaft while retrieving the device through the sheath and the balloon remained in the patient.The patient had surgery to remove the balloon.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: upon reaching the lesion site, the terumo angled glide 0.035x260cm wire was removed and replaced with a 0.014" nitrex guidewire.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: the physician was intending to use a rapidcross with a 6f 90cm terumo sheath, terumo angled glide wire 0.035x260cm to treat a soft plaque lesion in the left cephalic vein.Hepsaline contrast was used to inflate the device.Negative prep was not performed, the lesion was not predilated.During first inflation a longitudinal burst occurred.A snare was used to retrieve the bigger balloon portion, the remaining balloon fragments needed to be removed surgically, all balloon fragments were removed from the patient.If information is provided in the future, a supplemental report will be issued.
|
| |
|