| Catalog Number |
121703058 |
| Medical Device Problem Code |
Loss of Osseointegration (2408)
|
| Health Effect - Clinical Codes |
Pain (1994); Inadequate Osseointegration (2646); No Code Available (3191)
|
| Date of Event |
01/07/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
|
| |
|
Event or Problem Description
|
|
The patient had been happy with his hip "untll" recently he presented at (b)(6) with pain.The x ray showed a worn taper head junction.After initial consultation the patient was listed for revision surgery.Unfortunately this had to be brought forward as the taper appeared to wear out very quickly and so disassociated from the head leaving a sharp taper and mom debris in the joint space.Upon revision mom debris was seen and the hip "debried" successfully.The cemented elite plus stem pulled out from femur with ease.The cup was also removed and revision implants used to correct the deformity.An eto was used to access the femur and remove the cement.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|