• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION E SERIES DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION E SERIES DEFIBRILLATOR Back to Search Results
Model Number E SERIES
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to zoll medical corporation; the malfunction was duplicated and attributed to the lithium coin battery on the system board.The device operator's guide instructs users to replace the lithium battery on the device every 5 years.This device is approximately 9 years old.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
 
Event or Problem Description
Complainant alleged that during a routine shift check by a clinician, the device would not power-up.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
E SERIES DEFIBRILLATOR
Common Device Name
E SERIES
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
MDR Report Key8279454
Report Number1220908-2019-00136
Device Sequence Number1527890
Product Code MKJ
UDI-Device Identifier00847946010409
UDI-Public00847946010409
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K042007
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 01/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberE SERIES
Device Catalogue NumberE SERIES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/14/2019
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/07/2019
Initial Report FDA Received Date01/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-