Model Number 945-605
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Burn(s) (1757); Burn, Thermal (2530)
Date of Event 01/03/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Associated manufacturers report - 3002807968-2019-00002.
 
Event or Problem Description
According to the complaint received, a patient sensor burn from a tcpo2 sensor used with a tcm400 monitor was reported.A (b)(6) year-old female patient being treated for chronic refractory osteomyelitis was subjected to in-chamber treatment on (b)(6) 2019, where a tcm400 monitor was applied.On (b)(6) 2019 the patient came back from hyperbaric oxygen treatment and a wound was identified on her leg in the area where the sensor was placed.The patient was admitted to the wound care center where the wound currently is being treated.
 
Event or Problem Description
A 48 year-old female patient being treated for chronic refractory osteomyelitis was subjected to in-chamber treatment on (b)(6) 2019, where a tcm400 monitor was applied.On (b)(6) 2019 the patient came back from hyperbaric oxygen treatment and a wound was identified on her leg in the area of where the sensor was placed.The patient was admitted to the wound care center where the wound currently is being treated.According to the complaint received, a patient sensor burn from a tcm400 monitor was reported.A 48 year-old female patient being treated for chronic refractory osteomyelitis was subjected to in-chamber treatment on (b)(6) 2019, where a tcm400 monitor was applied.On (b)(6) 2019 the patient came back from hyperbaric oxygen treatment and a wound was identified on her leg in the area of where the sensor was placed.The patient was admitted to the wound care center where the wound currently is being treated.
 
Additional Manufacturer Narrative
The affected electrode was tested in a monitor and calibrated, hereafter it measured correctly.The tc-electrode will be sent to production for additional testing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TC ELECTRODE
Common Device Name
TC ELECTRODE
MDR Report Key8280937
Report Number1523456-2019-00001
Device Sequence Number1686221
Product Code LPP
UDI-Device Identifier05700699456052
UDI-Public(01)05700699456052(10)TQ01
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 06/26/2019
Report Date (Section F) 01/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device No Information
Device Model Number945-605
Device Catalogue Number945-605
Device Lot NumberTQ-01 027
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date01/07/2019
Device Age11 YR
Event Location Hospital
Date Report to Manufacturer01/08/2019
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date01/28/2019
Supplement Report FDA Received Date06/27/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device N
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Date Report Sent to FDA01/25/2019
-
-