DEPUY MITEK LLC US LUPINE LOOP PLUS ANCHOR W/OC; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
| Catalog Number |
210708 |
| Medical Device Problem Codes |
Deformation Due to Compressive Stress (2889); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/12/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If additional information should become available, a supplemental medwatch will be submitted accordingly.Udi: (b)(4).Investigation summary: the investigation is documented on the attached document "(b)(4)".The complaint devices were received and evaluated.Visual inspection confirms the reported issue of bent anchor.The anchor was bent at around 60-degree angle; however, the suture was still in place and intact upon the anchor and handle of the device.It was determined that a manufacturing engineering action ((b)(4)) would be opened to review the manufacturing process used to assemble these devices.The mia investigated the dhr of the devices and the lot was released in accordance to established procedures.In the assembly process, there is a 100% verification in the assembly step done by the operators.The maximum temperature reached during the vacuum dry process was 121.3°f (about 49.6°c).The fixtures used in the manufacturing of these devices have preventative maintenance performed every six months.Operators were informed about these complaints, and have been sensibilised to make more attention during the lupine assembly process.The manufacturing process was performed according to the procedures and is not the root cause of this issue.A definite root cause for this complaint cannot be confirmed.This file will remain receptive to any potential forthcoming information received that is pertinent and germane to this issue.A review into the depuy synthes mitek complaints system revealed two other similar complaints for this lot of (b)(4) devices that were released to distribution.At this point in time, no corrective action is required and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Device history lot: the batch review lot l320096 code product 210708 showed that the (b)(4) devices were conformed to in process and finished goods specifications when released to stock on march 13, 2017.This report is being filed from the (b)(4) as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
|
| |
|
Event or Problem Description
|
|
It was reported by the affiliate in (b)(6) that during an unspecified surgical procedure, it was observed that the lupine loop plus anchor w/oc device was defective.According to the reporter, after drilling, the device was opened and before insertion into the guide it was when the issue was observed.The surgery did not get hampered as extra implants were available and a new implant was opened and used.There was a delay of about one minute in the surgery.The surgeon has not raised any concern or complaint.There was patient involvement reported.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|