| Model Number |
9733449 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
01/23/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No patient information provided as no patient was involved in this concern.No parts have been received by the manufacturer for evaluation.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that, while outside of a procedure, the straight suction would verify intermittently.It was noted that the reported issue was confirmed with a demo head.There was no patient present when this issue was identified.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The suction was returned to the manufacturer for evaluation.Testing found that the instrument had difficulties verifying as the instrument returned a high divot error when attached to a known operational instrument tracker.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.If information is provided in the future, a supplemental report will be issued.
|
| |
|