• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION MUSTANG; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION MUSTANG; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number 24674
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/08/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The chronic totally occluded target lesion was located in the mildly tortuous and severely calcified superficial femoral artery.Two mustang balloon catheters (6.0 x 40,135cm and 6.0 x 60,135cm) were advanced for pre-dilatation.Both balloons ruptured during first inflation at 10 atmospheres nominal pressure for 30 seconds and contrast leakage was noted.The device was removed.Non-bsc devices also ruptured.A drug-eluting stent was deployed and the procedure completed.No patient serious injury or adverse events were reported and the patient status was fine.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: the device was returned for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.Blood was identified within the balloon and inflation lumen which is evidence of a device leak.A microscopic examination identified a longitudinal tear on the balloon material beginning approximately 11mm proximal of the distal markerband and extending approximately 17mm proximally across the balloon material.An examination of the balloon material and markerbands identified no issues which could potentially have contributed to this complaint.A visual and microscopic examination observed no issues with the tip of the device which could potentially have contributed to the complaint incident.A visual and tactile examination found no kinks or damage along the shaft of the device.No other issues were identified during the product analysis.
 
Event or Problem Description
It was reported that balloon rupture occurred.The chronic totally occluded target lesion was located in the mildly tortuous and severely calcified superficial femoral artery.Two mustang balloon catheters (6.0 x 40,135cm and 6.0 x 60,135cm) were advanced for pre-dilatation.Both balloons ruptured during first inflation at 10 atmospheres nominal pressure for 30 seconds and contrast leakage was noted.The device was removed.Non-bsc devices also ruptured.A drug-eluting stent was deployed and the procedure completed.No patient serious injury or adverse events were reported and the patient status was fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MUSTANG
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key8281523
Report Number2134265-2019-00301
Device Sequence Number11282824
Product Code FGE
UDI-Device Identifier08714729793939
UDI-Public08714729793939
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K141521
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date09/11/2021
Device Model Number24674
Device Catalogue Number24674
Device Lot Number0022639835
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/31/2019
Initial Date Received by Manufacturer 01/09/2019
Supplement Date Received by Manufacturer02/28/2019
Initial Report FDA Received Date01/28/2019
Supplement Report FDA Received Date03/19/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age71 YR
-
-