• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US MILAGRO ADVANCE SCREW 8X30MM; ORTHOPAEDIC BONE SCREW, BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US MILAGRO ADVANCE SCREW 8X30MM; ORTHOPAEDIC BONE SCREW, BIOABSORBABLE Back to Search Results
Catalog Number 231821
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/01/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If additional information should become available, a supplemental medwatch will be submitted accordingly.(b)(4).The exp date is currently unavailable.This report is being filed from the (b)(4) as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.The complaint device was received and evaluated.Visual observation confirms the screw is completely broke and there is a crack diagonally along the length of the screw.The threads of the screw were examined, and it was found to be distressed, which indicates that the failure happened during insertion.No other anomalies were found with the screw.Possible root causes could be that the patient had hard quality bone, and the user did not notch before inserting the screw.For hard bone, notching and tapping is required.This failure could be attributed to user technique issue.Further, a review into the depuy synthes mitek complaints system revealed no other complaints of any kind for this lot of devices that were released to distribution.At this point in time, no corrective action is required and no further action is warranted.This file will remain receptive to any potential forthcoming information received that is pertinent and germane to this issue.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.No nonconformances were identified for this part-lot number combination per qlik query executed 11/12/2018.Review conducted per (b)(4).
 
Event or Problem Description
This is report 2 of 2 for the same event.It was reported by the affiliate in (b)(6) that during an unspecified surgical procedure, it was observed that the milagro advance screw 8x30mm device broke during the screw-insertion over the guide-wire with the universal-screwdriver.It was an 7mm tibia-tunnel, with a short distal drill 8.5.The surgeon took the pieces out and insert a new one but it also broke.It was reported that the second one stayed in the patient as the screw broke near the screw-head.It was reported that the second screw was inserted into the same hole as well.There was no delay of greater than 30 minutes in the surgical procedure.There was patient involvement reported.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MILAGRO ADVANCE SCREW 8X30MM
Common Device Name
ORTHOPAEDIC BONE SCREW, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key8281914
Report Number1221934-2018-51685
Device Sequence Number19310660
Product Code HWC
UDI-Device Identifier10886705022267
UDI-Public10886705022267
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
K123362
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/01/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number231821
Device Lot Number3840999
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/30/2016
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/01/2016
Initial Report FDA Received Date01/28/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-