Model Number 7214
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/31/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Returned product consisted of a nc emerge balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed multiple kinks along the whole device.The hypotube is separated 49.1cm from the hub.Microscopic examination revealed no additional damages.There is blood present in the guidewire lumen and on the balloon folds.The balloon is still tightly folded.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis confirmed the hypotube is separated but appeared to be kinked prior to separation.
 
Event or Problem Description
It was reported that shaft break occurred.The target lesion was located in a saphenous vein graft to the obtuse marginal branch.While inserting a 3.00mm x 20mm nc emerge balloon catheter over the wire and into the guider, it was noted that the shaft broke in half.No patient complications were reported and the patient status is fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC EMERGE
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key8282150
Report Number2134265-2019-00362
Device Sequence Number19430502
Product Code LOX
UDI-Device Identifier08714729846659
UDI-Public08714729846659
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K141236
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/27/2020
Device Model Number7214
Device Catalogue Number7214
Device Lot Number0022720047
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/02/2019
Initial Report FDA Received Date01/28/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No