| Model Number |
7015 |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/15/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: the device was returned for analysis.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed kinks at 88.3cm, 96.5cm and 99.3cm from the hub.The hypotube is separated 103.1cm from the hub and appears to have been kinked prior to separation.Microscopic examination revealed no additional damages.There is contrast on the balloon folds and the balloon is still tightly folded.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 03 jan 2019.It was reported that crossing difficulties were encountered.The target lesion was located in the left anterior descending artery.A 3.25mm x 15mm quantum maverick balloon catheter was advanced but failed to cross the lesion.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.However, returned device analysis revealed shaft detached/separated.
|
| |
|