• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: US SURGICAL PUERTO RICO TA; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

US SURGICAL PUERTO RICO TA; STAPLE, IMPLANTABLE Back to Search Results
Model Number TA9048L
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 10/30/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The instrument was received loaded with a single use loading unit (sulu).Visual inspection of the sulu cartridge noted that a full staple complement was present.Functionally, the instrument and sulu were applied to test media with proper staple formation.It may occur when the user mistakes the tactile pre-clamp feature for the full firing stroke.This feature allows the handle to be released, without returning to its original position, for final positioning of the tissue.After the tissue is properly positioned, the handle stroke must be continued to full clamp position.Staples will only fire after the fully clamped handle returns to its original position and is squeezed a second time.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter during a colon resection a decision was made to perform a segmental l resection, rather than extend right hemicolectomy as discussed with the patient, the functional end to end anastomosis between the traverse colon with both device, however the devices did not fire and the anastomosis was sutured instead.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
MDR Report Key8285747
Report Number2647580-2019-00548
Device Sequence Number2701962
Product Code GDW
UDI-Device Identifier20884521043265
UDI-Public20884521043265
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K111825
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 01/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTA9048L
Device Catalogue NumberTA9048L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/10/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/07/2019
Initial Report FDA Received Date01/29/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-