(b)(4).Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The instrument was received loaded with a single use loading unit (sulu).Visual inspection of the sulu cartridge noted that a full staple complement was present.Functionally, the instrument and sulu were applied to test media with proper staple formation.It may occur when the user mistakes the tactile pre-clamp feature for the full firing stroke.This feature allows the handle to be released, without returning to its original position, for final positioning of the tissue.After the tissue is properly positioned, the handle stroke must be continued to full clamp position.Staples will only fire after the fully clamped handle returns to its original position and is squeezed a second time.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
|
|
According to the reporter during a colon resection a decision was made to perform a segmental l resection, rather than extend right hemicolectomy as discussed with the patient, the functional end to end anastomosis between the traverse colon with both device, however the devices did not fire and the anastomosis was sutured instead.
|