• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US TCK1 HD CAMERA HEAD; ENDOSCOPIC VIDEO CAMERA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US TCK1 HD CAMERA HEAD; ENDOSCOPIC VIDEO CAMERA Back to Search Results
Catalog Number 242400
Device Problem Operating System Becomes Nonfunctional (2996)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
If additional information should become available, a supplemental medwatch will be submitted accordingly.Udi: (b)(4).Performed service & repair.Nothing noted in service history review that could have contributed to the complaint.Attached box label, and service report form.Could not duplicate the customer's complaint that the image cuts in and out.Multiple attempts to duplicate the complaint were made, and the unit passed all attempts.The unit passes all functional and diagnostic testing and is fully operational.A non-conformance search was performed for this product code 242400, lot om01460 combination and no non-conformances were identified.No further information regarding the technique or instruments used has been provided to determine a root cause for this failure.If any additional information is obtained, this complaint will be re-opened to capture that information.At this point in time, no corrective action is required, and no further action is warranted.This report is being filed from the etq complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
 
Event Description
It was reported by the customer that during an unspecified surgical procedure, it was observed that the image on the tck1 hd camera head device was cutting in and out.There was no delay in the surgical procedure as an identical spare device was used to complete the procedure.There was patient involvement reported.There were no injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TCK1 HD CAMERA HEAD
Type of Device
ENDOSCOPIC VIDEO CAMERA
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6013142063
MDR Report Key8286648
MDR Text Key136855197
Report Number1221934-2018-53771
Device Sequence Number1
Product Code FWF
UDI-Device Identifier10886705026258
UDI-Public10886705026258
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/29/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number242400
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/26/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/09/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-