MEDTRONIC SOFAMOR DANEK USA, INC CD HORIZON BALLAST SPINAL SYSTEM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
|
Back to Search Results |
|
| Catalog Number |
25630018590 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Discomfort (2330)
|
| Date of Event |
11/27/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.X ray review results: post-op x-rays for thoraco-lumbar-sacral stabilization with pre revision x-ray.On the pre-revision x-rays there are fractured iliac screws bilaterally.Fusion status is unknown.The post revision x-ray, a fractured portion of the gearshift probe is seen in the iliac wing.This doesn't appeared to interfered with the placement of the revision screw.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was pre-operatively diagnosed with failed fusion and underwent screw placement surgery at s2- iliac levels.Post-operatively, the screws were broken at the shafts.As a result, revision surgery was performed to replace the product.The screw shaft of the broken product remained inside the patient.No patient symptoms or complications were reported as a result of this event.
|
| |
|